Fft Gating (Opt 238) - Keysight N9913A User Manual

Fieldfox analyzers
Hide thumbs Also See for N9913A:
Table of Contents

Advertisement

SA (Spectrum Analyzer) Mode
SA Mode Settings
Trigger Settings Annotation
Trigger settings are annotated on the FieldFox screen as highlighted in
the following image:
— Square: Trig Type (VidT), AutoTrigger (a), Pos Slope (/), Neg Delay (delay-)
— Horizontal arrow: Trig Level (green horizontal graticule at -20 dBm)
— Vertical arrow: Trig Position T-zero indicators (green arrows at 2nd vertical

FFT Gating (Opt 238)

Time-gated spectrum analysis allows you to obtain spectral information about
signals occupying the same part of the frequency spectrum that are separated
in the time domain. Using an external trigger signal to coordinate the
separation of these signals, you can perform the following operations:
— Measure any one of several signals separated in time; for example, you can
— Measure the spectrum of a signal in one time slot of a TDMA system.
— Exclude the spectrum of interfering signals, such as periodic pulse edge
172
——10: T-zero occurs at the far right graticule.
— Then press Enter.
graticule)
separate the spectra of two radios time-sharing a single frequency.
transients that exist for only a limited time.