RTSA (Real-Time Spectrum Analyzer) Mode—Option 350 (CPU2 Only)
Record/Playback (RTSA Option)

Record Source

RTSA can only record one of the four possible traces. The simplest way to use
Record Playback is to use First Active Trace (default trace).
For this discussion, it is important to understand Trace States. Learn more
in
Figure 10-7
FieldFox RTSA Mode: Data Flow and Record Playback Data Points
RTSA Data Recording:
Density Spectrum information is not recorded. Only the trace data is saved.
— Data is recorded at the trace state processing block.
Keysight N9927-90001 User's Guide
Record Interval
Sets the amount of time to wait between individual trace
recordings. Default is MIN, which is to wait 0 seconds (record traces as fast
as possible). Enter a value in seconds between 0 (no delay) and 10,000
seconds - about 3 hrs) using the numeric keypad,
knob.
Segment Count
Sets the number of traces to record, after which the
recording will automatically pause. Press Pause/Resume or Record to
capture another N traces, or Stop to end recording. Default is OFF, which is
NO limit to the number of traces to be recorded.
When set, a counter appears in the lower-left corner which counts UP to
the specified number of recordings.
"Trace Display States (SA Mode)" on page
arrows, or the rotary
175.