RTSA (Real-Time Spectrum Analyzer) Mode—Option 350 (CPU2 Only)
RTSA Mode Settings
Average Count
The Average Count setting is used mainly with the RMS Average Trace State
described above. (But, also applies to the MaxHold or MinHold trace states.) In
the RMS Average Trace State, the Average Count setting determines the
number of sweeps to average. The higher the average count, the greater the
amount of noise reduction.
For all three of these Trace States (RMS Average, MaxHold, or MinHold), when
Sweep 3
perform <n> sweeps, then return to Hold.
How to set Average Count
— Press Meas Setup 4
— Then
— Enter a value from 1 to 10,000 using the numeric keypad, the
— Press Enter
Alignments
The RTSA mode requires all three alignment types: amplitude, RF burst, and
channel equalization. Refer the SA chapter's
information on all Alignments.
Display Line
A display line is a simple, horizontal line that can be placed at any amplitude
level on the RTSA screen. Use a display line as mental guide for visual
feedback.
Keysight N9927-90001 User's Guide
Continuous
is set to OFF, press
Average Count
or the rotary knob.
Restart
to reset the sweep count to 1,
"Alignments" on page 177
arrows,
for
221