Marker Functions - Keysight N9913A User Manual

Fieldfox analyzers
Hide thumbs Also See for N9913A:
Table of Contents

Advertisement

Peak Excursion and Peak Threshold example—the settings for the trace in the following
image are:
——Threshold: –10 dB
——Excursion: 1 dB
——Scale = 1 dB / Division.
Peak A = Valid Peak (Above Threshold and Excursion Settings)
Peak B = Invalid Peak (Below 1 dB Excursion Setting)
Peak C = Invalid Peak (Below –10 dB Threshold Setting)

Marker Functions

Marker functions are used to change FieldFox frequency or distance settings to those of the active
marker.
How to use Marker Functions
— Activate the marker to move.
— Then press Mkr ->/Tools
— Then
Mkr ->
(this step not required for I/Q Analyzer mode)
Then choose one of the following:
Mkr -> Center
(Not available in DTF, I/Q Analyzer (waveform), or SA mode with zero span) The
center frequency of the measurement becomes the value of the active marker. The frequency
span is adjusted accordingly.
Mkr -> Delay
(NA mode ONLY.) The group delay value of the active marker becomes the
electrical delay setting of the trace. Learn how to set group delay format in
Settings" on page
91e. Learn more about electrical delay in
Mkr-> RefLev
(IQ Analyzer (NOT waveform) and SA mode ONLY) The acquisition becomes the
magnitude of the active marker.
Mkr->TuneFreq
(SA mode with Tune & Listen ONLY) The Tune & Listen Frequency becomes the
frequency of the active marker.
The following applies to DTF Measurements ONLY:
Keysight N9927-90001 User's Guide
Data Analysis Features
All about Markers
"Calibration
"Electrical Delay" on page
95.
459