Calibration for NA, CAT, and VVM Modes
— Press Meas Setup 4
— Then
— Then
— Under Transform Stimulus Settings, set the Start and Stop frequencies to
— Set Stimulus = Bandpass Impulse
— Press
— Press Meas Setup 4 or Back
— Then
— Set Media = Waveguide
— Set Cutoff Frequency. This is the absolute minimum frequency of the
Enter the calculated Effective VF value into the Velocity Factor setting.

Enhanced Response Optimization

When Enhanced Response Cal Type is selected, either for one calibration or for
CalReady, this setting optimizes the calibration based on the type of DUT
being measured. See also: CalReady Properties in
This setting does NOT survive Preset.
— Press Cal 5 then
— Then
— Then choose from:
Keysight N9927-90001 User's Guide
Transform
Transform Settings
those of the Waveguide.
Done
Calibration Settings
waveguide. This value must be less than the Start Frequency of the
Waveguide.
More
Enh.Response
NonReciprocal
(default) An amplifier is a Non-Reciprocal device
because it has gain in the forward direction, and very high loss
(isolation) in the reverse direction. This choice provides the best
correction for non-reciprocal devices, and reasonable correction for
reciprocal devices.
Reciprocal
A reciprocal DUT is a device in which the insertion loss
through the device is equal in both the forward (S21) and reverse
(S12) directions. A cable is a reciprocal device. This choice provides
the best correction for reciprocal devices. However, S11
measurements on non-reciprocal devices will appear to have more
return loss than the non-reciprocal choice.
"CalReady" on page
121.
137