General Specifications; Power Requirements - Agilent Technologies U3025AE10 User Manual

Hide thumbs Also See for U3025AE10:
Table of Contents

Advertisement

U3025AE10

General Specifications

Specifications for the Test Set are characteristic for the system performance of the PNA and
Test Set. Actual performance of the system is based on the customer's PNA and options that
are used with the Test Set. A functional certificate is only offered for the Test Set.
An N-Port calibration should be performed for optimum measurement accuracy.
A periodic calibration is not required. The Operators Check should be performed after
System Setup, or if performance is in question.
NOTE

Power Requirements

Verify that the required ac power is available before installing the Test Set to the PNA.
• 100/120/220/240 VAC (50/60Hz)
• The instruments can operate with mains supply voltage fluctuations up to ± 10% of
the nominal voltage.
• Air conditioning equipment (or other motor–operated equipment) should not be placed
on the same ac line that powers the Test Set and PNA.
Table 1
determine the electrical and cooling requirements.
Table 1
Standard Equipment
Instrument
U3025AE10
WARNING
6
This Manual:http://www.manuallib.com/agilent/u3025ae10-test-set-user-guide.html
When connected to a PNA, this Test Set will degrade the performance at the
test ports. The internal switch paths reduce test port power and power to the
receivers. This affects the test port power of the PNA and also reduces
dynamic range. The reflection tracking values measured in the
Check Procedures" on page 32
range to determine the approximate performance of the system.
contains the maximum wattage for all instruments. This table can be use to
Power Requirements
Maximum
Wattage
350
This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall be
inserted only into a socket outlet provided with a protective earth
contact. Any interruption of the protective conductor, inside or
outside the instrument, is likely to make the instrument dangerous.
Intentional interruption is prohibited.
can be subtracted from the analyzers dynamic
General Specifications
"Operational
User's Guide U3025-90001