Chapter 3
Meter Setup Menus
|
32
Orion Star A220 Series Portable Meter User Manual
|

Log View Menu

Use the Log View menu to access the data log and calibration log. All information in the data
and calibration logs is protected by the meter's nonvolatile memory, so stored information is
saved in the meter even when it is not connected to a power supply.

Data Log

Orion Star A220 series portable meters offer a 1000 point data log. Each point includes the
actively displayed measurements, depending on the meter model and display configuration,
with date and time stamp. When the data log function is turned on, the read type selected for
each displayed channel (Auto-Read or Continuous) determines how the point is saved to the
data log.
1. In the measurement mode, press the setup key.
2. Press the p, q, t or u key to highlight View Log and press the f3 (select) key.
3. Press the p or q key to highlight Data Log and press the f2 (select) key.
4. The meter will display a list of the data log points. The list shows the sequential number of
the points and the data and time the point was saved.
5. To view measurement information for an individual point:
a. Press the p or q key to highlight a point and then press the f2 (select) key.
b. Press the f1 (back) key to return to the list of data log points.
6. To export the data log to a printer or computer:
a. Press the f3 (options) key, press the p or q key to highlight Log Export and then
press the f2 (accept) key.
b. Press the p or q key to highlight Today, Last 10, Range or All and then press the f2
(accept) key.
i. If Range is selected, press the p or q key to highlight the number of the first
(from) and last (to) points to be exported, use the popup number entry screen
to edit the numbers and press the f2 (accept) key to save the entered range.
c. Press the f2 (yes) key to confirm that the meter is interfaced with a printer or computer.
The meter will automatically return to the data log list once the export is completed.
7. To clear the data log memory:
a. Press the f3 (options) key, press the p or q key to highlight Log Clear and then
press the f2 (accept) key.
b. Press the f2 (yes) key to confirm erasing the entire data log from the meter's memory.
c. Press the f3 (edit) key, use the popup number entry screen to enter the default meter
password of 111111 and then press the f2 (done) key.
d. Press the f2 (accept) key and wait for the data log to be cleared.