Specifications
Speed
Time
1-4
Description
• Set speed:
— 1,000 to rated speed in 100-rpm increments
• Maximum speed:
— XPN-80: 80,000 rpm
— XPN-90: 90,000 rpm
— XPN-100: 100,000 rpm
• Speed display:
— Actual measured rotor speed in 10-rpm increments below 1000 rpm
and 100-rpm increments ≥1000 rpm
• Speed control:
— Actual measured rotor speed, ± 2 rpm at steady state (1,000 rpm to
rated speed)
• Acceleration:
— 10 profiles
• Deceleration:
— 11 profiles, including coast
• Braking:
— Regenerative, power reclamation
• Set time:
— Up to 999 hours 59 minutes, including Hold runs
14
— Up to 3.94780 x 10
• Time display:
— Indicates time remaining in timed runs, time elapsed in Hold runs,
or estimated time remaining in ω²t runs
• Time accuracy:
— ± 70 ppm (6 seconds/day)
• Run modes:
— Time, ω²t, and RCF
radians squared per second in ω²t mode
PN B08115AF