Reduce Rotor Speed for Precipitating Solutions Page

Reduce Rotor Speed for Precipitating Solutions Page
Figure 8.65 Reduce Rotor Speed for Precipitating Solutions Page
To display the
button on the
Solutions
to avoid precipitation of CsCl during centrifugation using concentrated CsCl solutions.
In addition to the Header Bar and the Footer Bar, the Reduce Rotor Speed for Precipitating Solutions
page contains the following elements:
• The
Select Rotor and Labware
Select the rotor and labware to use for the calculation and select
Rotor Speed for Precipitating Solutions page.
• The
Average Density
the calculation. Enter the average density and select
to the Reduce Rotor Speed for Precipitating Solutions page.
• When you have entered the rotor and average density, the
field displays the calculated speed.
• The
Done
8-62
Reduce Rotor Speed for Precipitating Solutions
Calculations
Page. Use this page to calculate the reduced run speed required
button displays the
button displays a numeric input page to set the density of the sample for
button dismisses the page.
page, select the
For Precipitating
Select Rotor and Labware Page
to return to the Reduce
OK
to perform the calculation and return
OK
Maximum Allowable Speed (RPM)
(Catalog).
PN B08115AF