Specifications; Control Features - Beckman Coulter Optima XE Series User Manual

Preparative ultracentrifuges for non-in vitro diagnostic use
Table of Contents

Advertisement

Description

Specifications

Specifications
Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data,
without guarantee.

Control Features

Specifications
Speed
Time
2-4
Description
• Set speed:
— 1,000 to rated speed in 100-rpm increments
• Speed display:
— Actual measured rotor speed in 10-rpm increments below 1000 rpm
and 100-rpm increments ≥1000 rpm
• Speed control:
— Actual measured rotor speed, ± 2 rpm at steady state (1,000 rpm
to rated speed)
• Acceleration:
— 10 profiles
• Deceleration:
— 11 profiles, including coast
• Braking:
— Regenerative, power reclamation
• Set time:
— Up to 999 hours 59 minutes, including Hold runs
— Up to 3.94780 x 10^14 radians squared per second in
• Time display:
— Indicates time remaining in timed runs, time elapsed in Hold runs,
or estimated time remaining in
• Time accuracy:
— ± 70 ppm (6 seconds/day)
• Run modes:
ω
²t
— Time,
, and RCF
ω
²t
mode
ω
²t
runs