Specifications and Preinstallation Requirements
Specifications
Only values with tolerances or limits are guaranteed data. Values without tolerances are
informative data, without guarantee.
Control Features
Specifications
Speed
Time
Temperature
J326XP-IM-5AB
Description
• Setting range:
— 100 to 26 000 rpm (in 100-rpm increments below 10 000 rpm,
500-rpm increments above 10 000 rpm), or equivalent RCF (in
100 g increments)
• Elutriation speed setting range:
— 0 to 5 000 rpm (in 10-rpm or 10 g increments)
• Speed display:
— from 0 to 10 000 rpm, display indicates actual rotor speed 10 rpm;
from 10 000 to 26 000 rpm, indicates actual rotor speed 0.1%, or
equivalent RCF
• Setting range:
— to 180 minutes or continuous (hold)
• Time display:
— Indicates time remaining (timed run) or time elapsed (hold run)
• Setting range:
— –10 to +40°C (in 1°C increments)
• Accuracy:
— rotor temperature controlled to within 2°C of set temperature (after
a
equilibration)
• Ambient temperature range:
— 16 to 38° (60 to 100°F)
• Cooling fluid:
— refrigerant 404A (HFC)
b
CHAPTER 1