Diagnostics
Diagnostics/User Messages Chart
Table C.1 Diagnostics/User Messages Chart (Continued)
Message
D213 - Power Supply
D214 - Power Supply
D216 - AC Power Out
of Range
A217 - Alert: AC
Power Voltage Sag
D300 - Rotor Speed
A301 - Alert: Rotor
Speed Adjusted
A302 - Alert: Inertia
Calibration
D303 - Speed Signals
D304 - Speed Signals
D305 - Inertia
A400 - Alert: Vacuum
Calibration
C-4
Definition/Result
System +24V supply too high
(>26.4VDC) or System +24V
supply too low (<21.6VDC)
System +5V supply too high
(>5.5VDC)
AC Power Out of Range
AC voltage sagged below
180Vac or AC voltage/
frequency is out of range
during braking
Rotor speed exceeds
maximum rated speed
Rotor set speed adjusted
Inertia calibration failure
Overspeed timing signal is <9
counts/revolution or >47
counts/revolution, or
overspeed timing signal is
unstable
No tachometer signal
Inertia check failure
Vacuum calibration failure
Recommended Action
Call Beckman Coulter Field Service.
Call Beckman Coulter Field Service.
Check electrical power source.
None
Call Beckman Coulter Field Service.
None
Call Beckman Coulter Field Service.
1. Check the condition of the rotor overspeed
disk.
2. Make sure the rotor is installed properly.
3. Verify the rotor load is within limits
specified in the rotor manual.
NOTE
This condition will result in a 200-
minute delay before the diagnostic can be
cleared and the door opened. The power
must be left on, until the delay period is
completed.
1. Make sure the rotor is installed properly.
2. Check the condition of the rotor overspeed
disk.
3. Verify the rotor load is within limits
specified in the rotor manual.
4. If the problem persists, call Beckman
Coulter Field Service.
Be sure rotor is loaded properly.
Call Beckman Coulter Field Service.
PN B08115AF