Audible Sounds - Beckman Coulter Optima XPN User Manual

Preparative ultracentrifuges
Table of Contents

Advertisement

Specification
Outputs
Environmental
a. Normally, only non-conductive pollution occurs; occasionally, however, a temporary conductivity caused by condensation
must be expected.

Audible Sounds

The Optima XPN instrument makes an audible sound for the following events:
• Boot up
• Start of Run
• End of Run
• Diagnostics/Alert
• Vacuum low enough to open door
• Door open during Zonal or Continuous Flow mode
Sound volume can be regulated through the
through the System Options. See
For safety, the Door Open during Zonal or Continuous Flow mode sound cannot be changed or
muted. It will play every 5 seconds when the door could be open.
PN B08115AF
Description
• Power Factor:
— 0.99
• Nominal Output Voltage:
— 200/240
• Output Frequency:
— 50/60 Hz
• Max Output Current:
— 8.5A
• Max Fault Current:
— 20A
• RoHS compliant material:
— 100% of content
• Cooling system:
— Thermoelectric (no CFC gases used)
• Noise output (1m in front of instrument):
— <51 dBA (Sound pressure level)
— 7.0 sones (Zwicker Loudness)
• Maximum heat dissipation into room under steady-state conditions:
— 3400 Btu/hr (1.0 kW)
• Pollution degree:
a
— 2
• HEPA filter:
— Available
Set Sound
Custom Sounds
Page.
Page, and some sounds can be customized
1
Description
Specifications
1-7

Advertisement

Table of Contents
loading

Table of Contents