Specifications; Control Features - Beckman Coulter Avanti J-25 Instruction Manual

High-performance centrifuge
Table of Contents

Advertisement

SPECIFICATIONS

CONTROL FEATURES

CFC
Preinstallation Requirements
Only values with tolerances or limits are guaranteed data. Values
without tolerances are informative data, without guarantee.
Speed
Setting range . . . . . . . . . . . . . .0 to 25 000 rpm (in 10-rpm increments)
Elutriation speed setting range. . . . . . . . . . . 0 to 6 000 rpm (in 10-rpm
Speed display
. . . . . . . . . . . . . . . . . . . .
Time
Setting range . . . . . . .1 min to 99 hr 59 min,
Actual display . . . . . . . . . . . indicates run time remaining (timed run),
ω
t setting range . . . . . . up to 9.99 × 10
2
ω
2
t actual display . . . . . . . indicates accumulated centrifugal effect to
Temperature
Setting range . . . . . . . . . . . . . . . . . . –20 to +40°C (in 1°C increments)
Accuracy . . . . . . . . . . . . . . . . . . rotor temperature controlled to within
Ambient temperature range . . . . . . . . . . . . . 16 to 38°C (60 to 100°F)
Cooling fluid . . . . . . . . . . . . . . . . . . . . . . . . . Refrigerant 404A (HFC)
Acceleration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . maximum, slow, or timed
Deceleration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . maximum, slow, timed
* During transient conditions, such as acceleration and deceleration, rotor temperature may be
outside this range.
To reach temperatures above ambient, the centrifuge is dependent on the frictional heat
generated inside the chamber during operation. At low run speeds or low ambient tempera-
tures, the centrifuge may not be able to achieve some higher temperatures.
Specifications and
or equivalent RCF (in 100 × g increments)
or 10 × g increments)
from 0 to 10 000 rpm, indicates actual
rotor speed ±10 rpm
from 10 000 to 25 000 rpm, indicates actual
rotor speed ±0.1%, or equivalent rcf
ω
2
t, or continuous (hold)
ω
2
t, or elapsed time (hold run)
14
radians squared per second
3 significant digits (in exponential notation)
±2°C of set temperature (after equilibration)*
(1 to 10 min from 0 to 500 rpm)
(1 to 10 min from 500 to 0 rpm), or off
1