Finishing the transmitter connection (Industrial enclosure)

Connection check-up
1. Firmly tighten cable glands and insert blanking plugs in unused cable entries.
2. Check individual wire installation by tugging firmly.
3. Remove O-ring from lid.
4. Reinstate lid and screw in until mechanical stop. Wind back lid by one turn.
5. Mount O-ring by pulling it over the lid and tighten lid cover until you feel friction from the O-
ring on both sides. Wind lid further by one quarter of a turn to seal on the O-ring.
6. Reinstate and tighten lid lock screw.
7. Ensure that moisture does not penetrate the inside of electronics enclosure. Bend cables
downward immediately before cable glands.
SITRANS FS230
Compact Operating Instructions, 05/2021, A5E38755607-AC