Download Print this page

Tools Required; Torque Settings - Agilent Technologies E4404B Installation Notes

Spectrum analyzers and emc analyzers low frequency extension option ukb retrofit kit
Hide thumbs Also See for E4404B:

Advertisement

Tools Required

• T-10 TORX screwdriver
• T-15 TORX screwdriver
• Flat-blade screwdriver
• Sharp knife or razor blade
• Function Generator
• Synthesized Sweeper
• Digital Voltmeter
• Power Meter, Dual-Channel
• RF Power Sensor (2 required)
• Microwave Power Sensor
• 50 Ω Termination
• Power Splitter
• 20 dB Attenuator
• APC 3.5 (m) Cable Assembly
• APC 3.5 (m) to Type N (m) Adapter
• APC 3.5 (m) to Type N (f) Adapter
• BNC Tee (f,m,f)
• Dual Banana to BNC (f) Adapter

Torque Settings

To avoid potential RFI leakage and prevent connector damage, tighten to the following torque limits:
Item
SMA Connectors
3-mm, T-10 TORX screws
4-mm, T-15 TORX screws
Installation Note E4404-90007
Torque
In-lb
N-cm
8.5
95
9
102
21
236
5

Advertisement

loading

This manual is also suitable for:

E4405bE7403aE7404a