Channel Power Subsection

Center Freq
:MEASure:CHPower:CENTer <freq>
:MEASure:CHPower:CENTer?
Sets the center frequency of the display.
Int BW
:MEASure:CHPower:IBW <freq>
:MEASure:CHPower:IBW?
Specifies the integration bandwidth ranging from 100 Hz to
3 GHz to calculate the power.
Channel Span
:MEASure:CHPower:SPAN <freq>
:MEASure:CHPower:SPAN?
Sets the analyzer span for the channel power measurement.
Be sure the span is set between 1 and 10 times the
integration bandwidth.
There are also some remote commands which are irrelative
to the front- panel keys.
Channel Power and Density
:MEASure:CHPower?
This command returns scalar results of main channel power,
and power density.
:MEASure:CHPower:CHPower?
This command returns the value of the channel power in
amplitude units.
:MEASure:CHPower:DENSity?
This command returns the value of the channel power
density in amplitude units/Hz.