Figure 13. 300 Khz Nb H-Adcp Mounting Holes - Teledyne WORKHORSE Operation Manual

Horizontal h-adcp
Hide thumbs Also See for WORKHORSE:
Table of Contents

Advertisement

May 2015
The two M12 tapped holes on the top edge of the mounting plate are used to help handle the
system during production of the H-ADCP.
Do not use these holes to mount the H-ADCP or connect any other metal structures to the
H-ADCP.
Poor Mount
H-ADCP mount is using the two M5 tapped holes to connect to a
metal structure.
Page 24
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Figure 13.
300 kHz NB H-ADCP Mounting Holes
This can lead to
Galvanic Corrosion
Photos by Drs. P.E. Meijer / CEO Aqua Vision
WorkHorse H-ADCP Operation Manual
and destruction of the H-ADCP.
Good Mount
H-ADCP mount uses the three user mounting holes with isolation
bushings. Does not use the two M5 tapped holes.