Cq - Transmit Power; Table 18: Polled Mode Commands - Teledyne WORKHORSE Operation Manual

Horizontal h-adcp
Hide thumbs Also See for WORKHORSE:
Table of Contents

Advertisement

WorkHorse H-ADCP Operation Manual
Table 18:
Polled Mode Commands
Command
!
+
-
D
E
T
The commands are not echoed and they don't need to be followed by a CR/LF pair. These
commands are only available when CP = 1.
Enabling polled mode disables the battery saver feature. Do not enable this mode when running
from batteries.
CQ – Transmit Power
Purpose
Allows the transmit power to be adjusted.
Format
CQnnn
nnn = 0, or 1 to 255 (0 = Low, 1 to 255 = High)
Range
Default
CQ255
Recommended Setting. The default setting for this command is recommended for most applications.
Description
Allows the transmit power to be set high or low.
This command only affects 75 kHz Long Ranger systems.
Although other frequency H-ADCPs allow the CQ command to be set to values other than the default,
the CQ command has no effect and is not supported by PlanADCP.
Description
Execute a Break reset
Increment internal clock by 1 second
Decrement internal clock by 1 second
Dump the last ensemble
Print the current ensemble number
Print the current time
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
May 2015
Page 143