WorkHorse H-ADCP Operation Manual

PT6 - Receive Bandwidth

This test measure the receive bandwidth of the system. The bandwidth varies with system frequency and
the WB command setting.
>PT6
Receive Bandwidth:
Sample
bw
rate
expect
Bm1
307
120
results
PASS
The H-ADCP should be in non-moving water during this test to get valid test results.
Table 27:
PT6 Receive Bandwidth Nominal Values
Bandwidth setting
Broad
Narrow
Beam fails if <50% or >125% of nominal value.

PT7 - RSSI Bandwidth

This test checks the RSSI filter circuits are working. Values listed are the indicated RSSI sampled at 1-ms
intervals after a "listen" ping.
>PT7
RSSI Time Constant:
RSSI Filter Strobe 1 =
time
Bm1
Bm2
Bm3
msec
cnts
cnts
cnts
1
6
6
2
11
12
3
15
16
4
20
21
5
23
24
6
26
27
7
28
29
8
30
31
9
32
33
10
34
35
nom
43
43
result
PASS
PASS
>
bw
bw
bw
bw
Bm2
Bm3
Bm4
91
93
88
88 Khz
PASS
PASS
PASS
WB command
0
1
38400 Hz
Bm4
cnts
7
8
14
15
19
20
23
25
27
28
30
31
32
33
34
35
36
37
37
38
42
43
PASS
PASS
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
300 kHz
600 kHz
79
200
14
40
May 2015