Table 41: Percent-Good Data Format; Table 42: Status Data Format - Teledyne WORKHORSE Operation Manual

Horizontal h-adcp
Hide thumbs Also See for WORKHORSE:
Table of Contents

Advertisement

WorkHorse H-ADCP Operation Manual
FIELD #1 = 50, FIELD #2 = 5, FIELD #3 = 45, FIELD #4 = 0
FIELD 1 - Percentage of Partial Solutions - Shows percentage of successful velocity calculations (50%) re-
sulting in Partial Solutions. A Partial Solution is a solution that does not result in an error velocity. This
occurs because there is not enough information to compute the error velocity.
FIELD 2 - Percentage of transformations rejected - Shows percent of error velocity (5%) that was less than
the WE-command setting. The WE command has a default of 5000 mm/s. This large WE setting effective-
ly prevents the H-ADCP from rejecting data based on error velocity.
FIELD 3 - Percentage of No Solution Transformations - 45% of the velocity data were rejected because
there was insufficient information (velocity Data) to attempt a transformation.
FIELD 4 - Percentage of Full Solutions - 0% of the velocity data collected during the ensemble for range
cell 60 generated a Full Solution. A Full Solution indicates that an error velocity is available.
Table 41:
Percent-Good Data Format
Hex Digit
Binary Byte
Field
1-4
1,2
ID Code
5,6
3
Depth cell 1, Field 1
7,8
4
Depth cell 1, Field 2
9,10
5
Depth cell 1, Field 3
11,12
6
Depth cell 1, Field 4
13-1028
7-514
Depth cell 2 – 128 (if
used)
These fields contain information about the status and quality of H-ADCP data. A value of 0 means the
measurement was good. A value of 1 means the measurement was bad.
Table 42:
Status Data Format
Hex Digit
Binary Byte
Field
1-4
1,2
ID Code
5,6
3
Depth cell 1, Field
1
7,8
4
Depth cell 1, Field
2
9,10
5
Depth cell 1, Field
3
11,12
6
Depth cell 1, Field
4
13-1028
7-514
Depth cell 2 – 128
(if used)
Description
Stores the percent-good data identification word (00 04h).
Stores percent-good data for depth cell #1, field 1. See above.
Stores percent-good data for depth cell #1, field 2. See above.
Stores percent-good data for depth cell #1, field 3. See above.
Stores percent-good data for depth cell #1, field 4. See above.
These fields store percent-good data for depth cells 2 through 128 (depending on the
WN-command), following the same format as listed above for depth cell 1.
Description
Stores the status data identification word (00 05h).
Stores status data for depth cell #1, beam #1. See above.
Stores status data for depth cell #1, beam #2. See above.
Stores status data for depth cell #1, beam #3. See above.
Stores status data for depth cell #1, beam #4. See above.
These fields store status data for depth cells 2 through 128 (depending on the WN com-
mand) for all four beams. These fields follow the same format as listed above for depth
cell 1.