Table 13: H-Adcp Factory Defaults - Teledyne WORKHORSE Operation Manual

Horizontal h-adcp
Hide thumbs Also See for WORKHORSE:
Table of Contents

Advertisement

WorkHorse H-ADCP Operation Manual
Table 12:
H-ADCP Input Command Summary
Command
WAnnn
WBn
WCnnn
WDnnn nnn nnn
WEnnnn
WFnnnn
WIn
WJn
WLsss,eee
WNnnn
WPnnnn
WQn
WSnnnn [min, max]
WTnnnn
WUn
WVnnn
WWnnn
WYn,nnn
Table 13:
H-ADCP Factory Defaults
Command
BA
BB
BC
BE
BF
BI
BK
BL
BM
BP
BR
BX
BZ
CB
CF
CL
CM
CN
CP
CQ
DB
DW
EA
EB
EC
ED
EH
EP
ER
ES
ET
EX
Description
False target threshold maximum (0 to 255 counts)
Mode 1 Bandwidth Control (0 = Wide, 1 = Narrow)
Low correlation threshold (0 to 255 counts)
Data Out (Vel;Cor;Amp PG;St;P0 P1;P2;P3)
Error correlation threshold (0 to 5000 mm/s)
Blank after transmit (0 to 9999 cm)
Clip data past bottom (0 = Off, 1 = On)
Receiver gain select (0 = Low, 1 = High)
Water reference layer
Number of depth cells (1 to 128)
Pings per ensemble (0 to 16384)
Sample ambient sound (0 = Off, 1 = On)
Depth cell size, 20 to 800 (300kHz), 10 to 800 (600kHz)
Transmit length (0 to 3200 cm)
Ping weight (0 = Box, 1 = Triangle)
Ambiguity velocity (002 to 480 cm/s radial)
Mode 1 Pings before Mode 4 Re-acquire
WM 1: Bandwidth (0=WB, 1=NB), Amb. Vel.(cm/s)
300 kHz
30
320
220
1000
0
20
0
160,320,480
5
0
1
350
4
411
11110
1
0
1
0
255
611
0
+0
+0
1500
0
0
+0
+0
0
+2500
01110
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
600 kHz
30
160
220
1000
0
10
0
80,160,240
5
0
1
150
4
411
11110
1
0
1
0
255
611
0
+0
+0
1500
0
0
+0
+0
0
+2500
01110
May 2015
Page 121

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents