Figure 3-20

To Connect MCSMU Output

The MCSMU output connection example is shown in
For the connection of the MCSMU, the N1255A connection box is required. See
"To Connect MCSMU Connection Box" on page 3-26
the N1255A Input terminals. To connect the N1255A Output terminals to the
connectors as shown in
MCSMU Output Connection
High Force
From N1255A Output
High Sense
Insulator
Plate
Low Force
From N1255A Output
Low Sense
Triaxial connector
The High Force/High Sense/Low Force/Low Sense lines should be connected to the
DUT as shown above. Wiring shown in
can be used. To simplify the connections, omit the wiring of the Common.
For the non-Kelvin connection, use the High Force and the Low Force only. Omit
the wiring of the High Sense and the Low Sense.
Keysight B1500A User's Guide, Ed ition 14
Figure
3-20, use the 16494A triaxial cable.
High Force
Common
High Sense
Low Force
Common
Low Sense
"To Connect SMU Output" on page 3-35
Mounting Connectors
Figure
3-20.
to connect the MCSMU to
To DUT
High terminal
To DUT
Low terminal
Coaxial cable
Wire
Installation
3- 41