Keysight Technologies B1500A User Manual page 181

Semiconductor device analyzer
Table of Contents

Advertisement

Table 4-23
Voltage applied to DUT
Applied voltage condition
Voltage measurement condition
a. Recommended minimum pulse width = settling time + recommended minimum measurement
window.
b. The time until the measured value settles to within ± 0.6 % of the final result value after the out-
put voltage is changed from the initial value (0 V). Minimum rise/fall time of 70 ns for 10 V, or
30 ns for 5 V is recommended for minimizing overshoot.
c. RMS noise measured over the recommended minimum measurement window.
1. Measurement conditions: The DUT is a resistive load between 1 k and 10 M. The capaci-
tance of the cable between the RSU and the DUT is 20 pF. Voltage is applied to the DUT by a
channel of WGFMU/RSU, and voltage measurement is performed by the same channel. (PG
mode for 5 V, Fast IV mode for 10 V)
Minimum timing parameters for voltage measurement
Recommended minimum pulse width
Measurement Range
Recommended minimum measurement window
Settling time
Noise (rms)
Keysight B1500A User's Guide, Ed ition 14
b
c
Specifications
1
5 V
105 ns
a
5 V
20 ns
85 ns
1.4 mV
Hardware
10 V
130 ns
10 V
20 ns
110 ns
1.4 mV
4- 41