Ground Unit (GNDU)
The GNDU is furnished standard with the mainframe.
Output voltage: 0 V ± 100 V
Maximum sink current: ± 4.2 A
Output terminal/connection: Triaxial connector, Kelvin (remote sensing)
Load capacitance: 1 F, supplemental characteristics
Cable resistance: supplemental characteristics
Terminal
Sense
Force
Keysight B1500A User's Guide, Ed ition 14
Cable resistance
 10 
< 1  (Is  1.6 A)
< 0.7  (1.6 A < Is  2.0 A)
< 0.35  (2.0 A < Is  4.2 A)
where Is is the current being sunk by the GNDU.