Ground Unit (GNDU)
The GNDU is furnished standard with the mainframe.
Output voltage: 0 V ± 100 V
•
•
Maximum sink current: ± 4.2 A
•
Output terminal/connection: Triaxial connector, Kelvin (remote sensing)
Load capacitance: 1 F, supplemental characteristics
•
•
Cable resistance: supplemental characteristics
Terminal
Sense
Force
Keysight B1500A User's Guide, Ed ition 14
Cable resistance
10
< 1 (Is 1.6 A)
< 0.7 (1.6 A < Is 2.0 A)
< 0.35 (2.0 A < Is 4.2 A)
where Is is the current being sunk by the GNDU.