Keysight Technologies B1500A User Manual page 132

Semiconductor device analyzer
Table of Contents

Advertisement

Installation
Capacitance Compensation When Using Switching Matrix
Table 3-9
Compensation Coefficients and Modifications
Compensation
3- 54
For E5250A:
# E5250A
C Compensation coefficient data table
#
#
CAUTION : Do not add or delete "REVISION" line and "DATAxx" line.
#
Change the value for R,L,C of DATA05,06,07 or 08.
#
REVISION
A.03.00
#
R [ohm]
DATA00
74.65E-3
DATA01
75.41E-3
DATA02
231.41E-3
DATA03
177.56E-3
DATA04
100.70E-3
DATA05
100.70E-3
DATA06
100.70E-3
DATA07
114.00E-3
DATA08
114.00E-3
DATA09
0.00E-3
# END of Data
The lines DATA05 to DATA08 should be modified for each measurement
environment. See
Table
coefficients
C2H
For the connector plate, change the R, L, C values in the lines.
(DATA05)
The value must be changed to the R, L, C values of the
C2H (DATA05) path and the C2L (DATA06) path (triaxial
C2L
cable with connector plate) shown in
(DATA06)
For Keysight B2220A probe card interface, do not modify the
lines.
C3H
Change the R, L, C values in the lines. The value must be
(DATA07)
changed to the R, L, C values of the C3H (DATA07) path and
the C3L (DATA08) path shown in
C3L
(DATA08)
For the connector plate, the coaxial cable with positioner will
be the appropriate path.
For Keysight B2220A probe card interface, the probe card will
be the appropriate path.
L [H]
C [F]
140.00E-9
58.44E-12
90.00E-9
67.13E-12
450.00E-9
178.85E-12
390.00E-9
135.45E-12
400.00E-9
80.00E-12
400.00E-9
80.00E-12
400.00E-9
80.00E-12
544.00E-9
130.00E-12
544.00E-9
130.00E-12
0.00E-9
1.20E-12
3-9. Do not modify the other lines.
Modifications of data file
Figure
Keysight B1500A User's Guide, Ed ition 14
# Frame Path 1
# Frame Path 2
# Card Path High
# Card Path Low
# Triax Cable
[/m]
# User Triax Cbl H [/m]
# User Triax Cbl L [/m]
# User Coax Cbl H
[/m]
# User Coax Cbl L
[/m]
# Stray Capacitance
Figure
3-22.
3-22.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents