Installation
Installing Accessories

To Connect 16442B Test Fixture (B1500A-A5F)

See this section if you use Keysight 16442B test fixture (B1500A-A5F). This
section describes how to set up the 16442B and how to connect the B1500A to the
16442B.
To Set up 16442B (B1500A-A5F)
You can stabilize the 16442B (B1500A-A5F) as shown in the figure below.
1. Put a stabilizer on both sides of the test fixture.
2. Attach each stabilizer to the test fixture with a flathead screw.
To Connect 16442B (B1500A-A5F)
Use the cables shown in
Table 3-2
to connect from the B1500A connectors to the
16442B's relative connectors.
Making non-Kelvin connection
NOTE
The Force terminals can be used to force and measure DC voltage or current. If you
want to simplify the cable connections, open the Sense terminals and connect the
Force terminals only to the test fixture by using the triaxial cables.
If you make the Kelvin connection (remote sensing), use both Force and Sense
terminals. Connecting the Force and Sense lines together at the terminal of the
device under test minimizes the measurement error caused by the residual resistance
of the connection cables. The Kelvin connection is effective for the low resistance
measurement and the high current measurement.
Keysight B1500A User's Guide, Ed ition 14
3- 13