Keysight Technologies B1500A User Manual page 62

Semiconductor device analyzer
Table of Contents

Advertisement

Table 2-1
Table 2-2
Introduction
Measurement Modules
HPSMU Output Voltage Ranges and Resolutions
Range
Output Value
0  |V|  2 V
2 V
0  |V|  20 V
20 V
0  |V|  40 V
40 V
0  |V|  100 V
100 V
0  |V|  200 V
200 V
HPSMU Output Current Ranges and Resolutions
Range
Output Value
0  |I|  1.15 nA
1 nA
0  |I|  11.5 nA
10 nA
0  |I|  115 nA
100 nA
1 A
0  |I|  1.15 A
10 A
0  |I|  11.5 A
100 A
0  |I|  115 A
0  |I|  1.15 mA
1 mA
0  |I|  11.5 mA
10 mA
0  |I|  50 mA
100 mA
50 mA < |I|  115 mA
0  |I|  50 mA
1 A
50 mA < |I|  125 mA
125 mA < |I|  500 mA
500 mA < |I|  1 A
2- 20
Setting
Resolution
100 V
1 mV
2 mV
5 mV
10 mV
Setting
Resolution
50 fA
500 fA
5 pA
50 pA
500 pA
5 nA
50 nA
500 nA
5 A
5 A
50 A
50 A
50 A
50 A
Keysight B1500A User's Guide, Ed ition 14
Maximum
Current
± 1000 mA
± 1000 mA
± 500 mA
± 125 mA
± 50 mA
Maximum
Voltage
± 200 V
± 200 V
± 200 V
± 200 V
± 200 V
± 200 V
± 200 V
± 200 V
± 200 V
± 100 V
± 200 V
± 100 V
± 40 V
± 20 V

Hide quick links: