To Interconnect Spgus - Keysight Technologies B1500A User Manual

Semiconductor device analyzer
Table of Contents

Advertisement

Figure 3-24
CAUTION
NOTE
Installation
About Plug-in Modules

To Interconnect SPGUs

If multiple SPGUs have been installed, connect the SPGUs by using the following
parts furnished with the SPGU. See
SPGU Sync cable, 16493Q-001, (number of SPGUs -1)2 ea.
RF connector cap, 1253-7431, 4 ea.
SPGU Module Connection Example, Four Modules
Slave SPGU
Master SPGU
If multiple SPGUs have been installed, connect the SPGUs as shown in
This connection makes it possible to perform the synchronous output by the
multiple SPGU channels. In this figure, the master SPGU is the module which
outputs the reference signal and the slave SPGUs are the modules which follow to
the reference signal.
Connect the RF connector cap to the Ref In and Sync In connectors on the master
SPGU and the Ref Out and Sync Out connectors on the last slave SPGU.
The Trig Out terminal must be connected to the trigger input terminal of the external
equipment to make the synchronized operation of it.
Connect the Ref Out/In, Sync Out/In, and Trig Out terminals to the specified
terminal properly. Connecting to the other terminal may result in damage to the
SPGU.
The B1500A does not support the synchronized operation between SPGU and
WGFMU.
3- 60
Figure
3-24.
RF connector cap
Ref In
Trig Out
Ref Out
Keysight B1500A User's Guide, Ed ition 14
SPGU Sync cable
Output 2
Sync In
Output 1
Sync Out
Figure
3-24.