Download Print this page
Keysight Technologies B2901A Demo Manual

Keysight Technologies B2901A Demo Manual

Making current-voltage measurement using smu, precision source/measure unit
Hide thumbs Also See for B2901A:

Advertisement

Quick Links

D E M O G U I D E
Making Current-Voltage
Measurement Using SMU
B2901A/02A/11A/12A Precision Source/Measure Unit

Advertisement

loading

Summary of Contents for Keysight Technologies B2901A

  • Page 1 D E M O G U I D E Making Current-Voltage Measurement Using SMU B2901A/02A/11A/12A Precision Source/Measure Unit...
  • Page 2 (IV) measurements more accurately and quickly than ever before. In addition, the B2901A/02A/11A/12A comes with an intuitive graphical user interface (GUI) and PC-based application software that make it easy for you to begin making productive measurements immediately.
  • Page 3 Measurement concept – Sourcing voltage to turn the LED on. – Limit value feature prevent the instrument from sourcing too much even if Source value is increased. – Stepping voltages from start voltage to stop voltage with making current measurement at each voltage step. –...
  • Page 4: Procedure Overview

    LAB 1: Turn on the LED and Measure LED Current Demonstration Objective 1. Change View mode to Single View This demo shows how easily a current-voltage measurement repeatedly until Single View a. Press can be made with the B2900A for Channel 1 is shown in the display. Series through the Light Emitting Diode (LED) characteristics 2.
  • Page 5 LAB 1: Turn on the LED and Measure LED Current (continued) d. Press and rotate until Source value achieves 3 V. You will see the measurement parameters turn orange, which means they reach Limit value. Measurement parameters (1) Press Source and rotate will turn orange until Source value achieves 3 V until Source value achieves 2 V and then press...
  • Page 6 LAB 1: Turn on the LED and Measure LED Current (continued) c. Press to stop making measurements periodically. “Auto” indicator is turned off 4. Turn off the channel output a. Press to edit Channel 1 Source value, and then enter 0 V to set Source value to 0 V.
  • Page 7 Lab 2: Make Current – Voltage (I – V) Sweep Measurement to get LED I – V Characteristics Demonstration Objective 1. Change View mode to Single View This demo shows how easily current-voltage characteristics repeatedly until Single View for a. Press can be obtained with the B2900A Channel 1 is shown in the display.
  • Page 8 Lab 2: Make Current – Voltage (I – V) Sweep Measurement to get LED I – V Characteristics to show Sweep Sub-Panel. d. Press , and then press (1) Press More… (2) Press Show Sweep e. Press , then press to turn on Single Linear Sweep Mode.
  • Page 9 Lab 2: Make Current – Voltage (I – V) Sweep Measurement to get LED I – V Characteristics 4. Performing the measurement a. Press Ch1 to turn on Channel 1 Output relay, and then press to perform a sweep measurement. During the measurement, the status information will show 5.
  • Page 10 Lab 2: Make Current – Voltage (I – V) Sweep Measurement to get LED I – V Characteristics 7. View the list of the measurement data a. If you’d like to see the list of the measurement result, press, then press to open Measure Result dialogue.
  • Page 11 Keysight office. The complete list is available at: www.keysight.com/find/contactus Page 11 Find us at www.keysight.com This information is subject to change without notice. © Keysight Technologies, 2018 - 2020, Published in USA , January 6, 2020, 5991-4248EN...

This manual is also suitable for:

B2912aB2902aB2911a