Specifications
Hardware
Ultra-fast IV measurement capability for the pulsed IV and transient IV such as
NBTI/PBTI, RTN, etc.
Waveform generation with 10 ns programmable resolution
Simultaneous high-speed IV measurement capability (200 MSa/s, 5 ns sampling
rate)
10 V peak-to-peak output
No load line effect accurate pulsed IV measurement by dynamic SMU
technology
Maximum Module Configuration
The B1500A can contain any combination of the following module. Then the total
power consumption of all SMU modules cannot exceed 84 W.
Up to 10 MPSMU
Up to 10 HRSMU
Up to 4 HPSMU
Up to 4 MCSMU
One MFCMU
Up to 5 HV-SPGU
Up to 5 WGFMU
When one or more WGFMU modules are installed in the B1500A mainframe, the
following table applies. Multiply the values given below by the number of installed
modules of that type and add the products together. The sum of the products must be
less than or equal to 59 for the configuration to be permissible.
WGFMU
Maximum Voltage between Common and Ground
 ± 42 V
4- 6
2
2
14
5
7
12
10