Keysight Technologies B1500A User Manual page 123

Semiconductor device analyzer
Table of Contents

Advertisement

Connections for High Current Measurements
When you force or measure a large current, you may want to use a Kelvin
connection (4-wire, remote sensing) to eliminate the residual resistance effects of
test leads and contacts. For example, you can use the following connections as
Kelvin connections on the test fixture.
Examples: Kelvin Connection
1
Force
Sense
Guard
Guard
To cancel the effects of the residual resistance, test leads must be connected as close
as possible to the DUT.
Keysight B1500A User's Guide, Ed ition 14
Connecting Measurement Devices
SMU(±200V= 1A Max)
2
Force
Sense
Guard
Guard
Drain
Gate
Substrate
Source
Installation
3 (Selected)
Force
Sense
Guard
Guard
3- 45

Hide quick links: