To install ASU
To connect
B1500A
NOTE
Installation
Installing Accessories
1. Fix the ASU in the shielding box. The ASU must be fixed to the best position for
accessing its connectors. Keysight N1254A-108 Magnetic Stand will be useful
for fixing the ASU.
2. Pass the 16493M (D-sub cable, triaxial cable), N1300A CMU cable, 16048D/E
test leads, BNC cable and so on through the cable hole of the 16495K. They will
be connected between the ASU and the instruments
3. Connect the D-sub cable to the ASU D-sub connector.
Connect the triaxial cable to the ASU Force terminal.
Connect the cable to the ASU CMU-cur/AUX In terminal. See
For the MFCMU, connect the N1300A CMU cable.
For Keysight 4284A/E4980A LCR meter, connect Keysight 16048D/E test
leads.
For the instrument other than the MFCMU or the 4284A/E4980A, connect
the BNC cable to the CMU-cur/AUX In terminal. And open the CMU-pot
terminal or cover it by using the BNC open cap.
4. Adjust the cable length in the shielding box, and catch in the cables by using the
cover of the 16495K.
5. Connect the cable from the DUT interface to the ASU output terminal.
For the non Kelvin connection, open the Sense terminal or cover it by using the
triaxial open cap.
1. Turn the B1500A off.
2. Connect the D-sub cable from the ASU to the SMU D-sub connector.
3. Connect the triaxial cable from the ASU to the SMU Force terminal.
To use the ASU, it must be connected to the SMU which is connected together when
the calibration is performed. The specifications are satisfied and guaranteed in this
condition only. So confirm the serial number of the ASU and connect it to the
dedicated SMU properly. See Keysight EasyEXPERT User's Guide to confirm the
serial number.
3- 18
Keysight B1500A User's Guide, Ed ition 14