5.2.1 Adapting to the cable length
The way the torque flange operates is dependent on the method used for
transmission between the rotor and the stator, and the following:
• Installation situation (e.g. cover, area free of metal)
• Length of cable
• Excitation voltage supply tolerances
For adaptation to different ratios, there are three switches in the stator
housing which are accessible when the stator cover is removed.
Switches
Fig. 5.1: Switches in the stator housing
Switch-
position
1
2
3
Please ensure that after switching to position 3, the calibration signal is not
initiated.
B 23.T10F.90 en
SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com
Typical application
a) Older amplifiers
b) For when the calibration signal is
unintentionally initiated with very short cable
Normal position (ex-works position)
For cable lengths over approx. 20 m