20
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com
3. Now refit the antenna segment removed under point 1. onto the stator
using two hexagon–socket screws and the serrated locking washers. Make
sure that for a defined transition resistance all required serrated locking
washers (see Fig. 4.2), die f are located under the nuts. Do not fully tighten
the screws yet.
4. Fit the upper screws for connecting the two antenna segments so that the
antenna ring is closed. Remember the serrated locking washers in this
case, too.
5. Now, fasten all antenna-segment screwed connections with a tightening
torque of 5 N⋅m.
6. Align the antenna and rotor so that the antenna encloses the rotor more or
less coaxially. Please adhere to the permitted alignment tolerances
specified in the technical data.
7. Fasten the stator-housing screwed connection.
8. Make sure that the opening in the lower
antenna-segment section is free from electroconductive
foreign substances.
CAUTION
In order to ensure proper functioning, the serrated locking washers (A5,
1–EST Din 6796 ZN/galvanised) must be renewed after the antenna
screws have been loosened three times.
HBM
SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com
T10F
Opening
B 23. T10F.90 en