4.4 Installing the rotor

CAUTION
For regular operation observe the mounting dimensions in all cases (in
particular area of metal parts, see page 46) .
Additional mounting instructions for the speed measuring system can be
found in chapter 4.7, page 23.
NOTE
In general, the rotor type-plate will no longer be visible after installation.
Therefore, additional adhesive labels including the major characteristics
have been supplied. They can be fixed on the stator or other relevant
test-rig components thus enabling required data such as e.g. the
calibration signal to be read off at any time.
1. Prior to installation, clean the plan surfaces on the torque flange and
counter flange. For safe torque transmission, the surfaces must be clean
and free from grease. Use a piece of cloth or paper soaked with a solvent.
Make sure that no solvent drips into the torque flange's interior and that the
transmitter coils do not suffer damage.
Measuring body
Fig. 4.1: Rotor screwed connection
HBM
SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com
8x DIN 912 hexagon–socket screw (Z) (10.9/12.9)
8x fastening screw (10.9, 12.9); observe
maximum thread depth Y!
T10F
B 23. T10F.90 en