8.1

Shielding design

The Greenline HBM shielding design ensures that the entire measurement
chain is completely enclosed in a Faraday cage, due to the special way the
cable shield is arranged (see also http://www.hbm.com/Greenline).
8.2

Signal ground

All devices - transducers, amplifiers and display devices - are located on an
earth potential (if necessary wire to a potential equalization line). If this is not
possible, the transducer should be fitted earth‐isolated.
WA...
A00546_11_YI0_01 HBM: public
Interference effects
23