Fitting Emergency Ventilation Bag; Connecting The Anaesthetic Gas Scavenging System Ags - Dräger Julian Instructions For Use Manual

Anaesthetic workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Fitting emergency ventilation bag

Example: Dräger Resutator 2000
Insert the adapter in the bag so that the rubber flange
is uniformly located in the groove of the adapter.
Press the retaining ring together completely with the
large eyelet pointing towards the bag and clamp it
over the bead on the bag.
1 Place the diaphragm in the bottom part of the valve
with the lip facing downwards.
Screw the upper part of the valve onto the bottom
part.
2 Plug the patient valve into the bag adapter.
Connecting the anaesthetic gas scavenging system
AGS
3 Connect the transfer hose to the excess gas
connector and to the port of the scavenging system.
4 Connect the suction hose to the port of the
scavenging system.
5 Connect the anaesthetic waste gas connector to the
suction hose.
6 Ensure that the second port on the scavenging
system is sealed with the sealing screw.
Note the Instructions for Use of the anaesthetic gas
scavenging system AGS.
1
2
3
3
6
3
4
Care
Assembling
5
111