Configuring in Standby Mode
Setting default values
Setting the measuring parameters
Select the line »parameters« with the rotary knob and
press to confirm.
The »parameters« menu appears in the left-hand field.
Display (example):
Switch SpO
measurement (optional) on/off:
2
Select the line »SpO
measurement« with the rotary
2
knob and press to confirm.
Select »on« or »off« with the rotary knob and press
to confirm.
Select the sample rate for gas measurement:
Select the line »sample rate« with the rotary knob
and press to confirm.
Select »60« or »200 mL/min« with the rotary knob
and press to confirm.
For paediatric use:
Select a sample rate of 200 mL/min in order to obtain
more accurate values on account of the high frequen-
cies and time constant of the measuring system.
Recirculate the measured gas, see page 109.
If the measured gas is not recirculated:
Increase the minute volume accordingly in order to
maintain the volume balance.
For adult use:
Select a sample rate of 60 or 200 mL/min.
Recommendation: select 200 mL/min and recirculate
the measured gas, see page 109.
Set CO
unit:
2
Select the line »CO
unit« with the rotary knob and
2
press to confirm. The last value selected is highlighted
by a grey background.
Select the required unit of measure by turning the
rotary knob and press to confirm.
84
Standby / Configuration
anesth. agent
calibration
SpO
-measure.
on
off
2
sample rate
60
200
mL/min
CO
unit
mmHg
kPa vol%
2
Menu for configuration of measurement parameters.
default settings
pulse tone
0 1 2 3 4
5
6 7 8 9
alarm sound
1 2 3
4
5 6 7 8 9
scale
adult
child
parameters
record
interfaces
alarm limits
curves
basic configurations
IPPV default settings
PCV default settings
fresh gas delivery