Adjustment Range For Default Alarm Limits - Dräger Julian Instructions For Use Manual

Anaesthetic workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Adjustment range for default alarm limits

Alarm limit
Adjustment range
SpO
51 to 100
2
_
[%]
50 to 99
H
Pulse
31 to 300
_
[1/min]
30 to 299
etCO
1 to 75
2
_
[mmHg]
0 to 74
MV
0.1 to 39
_
[L/min]
0 to 38.9
FiO
19 to 100
2
_
[Vol.%]
18 to 99
Fi Hal.
0.1 to 7
_
[Vol.%]
0 to 6.9
Fi Iso.
0.1 to 7
_
[Vol.%]
0 to 6.9
Fi Enf.
0.1 to 7
_
[Vol.%]
0 to 6.9
Fi Des.
0.1 to 21.9
_
[Vol.%]
0 to 21.8
Fi Sev.
_
0.1 to 9.9
[Vol.%]
0 to 9.8
PAW
5 to 98
_
[mbar]
0 to 35
Fixed alarm limits / times:
FiCO
Fixed upper alarm limit 5 mmHg
2
(deactivated in MAN/SPONT).
Apnoea pressure
after 15 seconds
Apnoea flow
after 15 seconds
Apnoea CO
after 15 seconds
2
(after 60 seconds in MAN/SPONT)
– – : The default value set on delivery is outside the
monitoring range and the corresponding alarm limit
is therefore switched off.
The newly set default alarm limits are valid whenever
the apparatus is switched on or started up from
standby mode.
Certain alarm limits are switched off automatically in
MAN/SPONT ventilation mode and when using the
external fresh gas outlet:
see tables, alarm limits for MAN/SPONT, page 39
and alarm limits for non-rebreathing systems, page 51.
Default value
set on delivery
– –
92
120
50
50
(mmHg)
– –
12
3.0
– –
20
1.5
– –
2.3
– –
3.4
– –
12.0
– –
3.4
– –
40
8
Configuring in Standby Mode
Setting default values
87