Record

This function is used to determine the type of event that
triggers an entry in the record list or a printout on the
connected record printer.
Select the line »record« with the rotary knob and
press to confirm.
The »record« menu appears on the left-hand side.
Display (example):

Time

interval-started (min)
A list entry is activated after a fixed time interval in
minutes.
NiBP-started
An entry is made after every NiBP measurement with new
measured values. This entry can be read by an external
Dräger monitor via the MEDIBUS interface.
Warning-started
An entry is made whenever an alarm is triggered.
Caution-started
An entry is made whenever a caution message is
triggered.
Turn the rotary knob to select the event triggering the
entry and press to confirm.

Configuring the interfaces

Julian has three serial interfaces:
COM1, COM2, COM3.
COM2 and COM3 are configured as MEDIBUS*
interfaces and COM1 as printer interface. The interfaces
can be adapted to the connected equipment.
Turn the rotary knob to select the line »interfaces«
and press to confirm.
The »interfaces« menu appears in the left-hand field.
Display (example):
COM2 is configured in the upper half of the menu and
COM1 and COM3 in the bottom half.
_____________
* MEDIBUS: Dräger communications protocol for medical equipment.
Standby / Configuration
anesth. agent
calibration
list entry:
time
intervall (min)
1
2
5
10
off
NiBP started
on
off
on
off
warning started
caution started
on
off
printer entry:
time
intervall (min)
1
2
5
10
off
NiBP started
on
off
warning started
on
off
on
caution started
off
Menu for configuration of list and printer entries.
Standby / Configuration
anesth. agent
calibration
COM 2 (Medibus):
baud rate
1.2
9.6
even
parity
8
data bits
1
stop bits
COM 1 (printer), COM 3 (Medibus):
config. of:
COM 1
COM 3
baude rate
1.2
2.4 4.8 9.6 19.2
parity
odd
even
none
data bits
7
8
stop bits
1
2
Menu for configuration of Medibus and printer ports.
Configuring in Standby Mode
Setting default values
default settings
pulse tone
0 1 2 3 4
5
6 7 8 9
alarm sound
1 2 3
4
5 6 7 8 9
scale
adult
child
parameters
record
interfaces
alarm limits
curves
basic configurations
IPPV default settings
PCV default settings
fresh gas delivery
default settings
pulse tone
0 1 2 3 4
5
6 7 8 9
alarm sound
1 2 3
4
5 6 7 8 9
scale
adult
child
parameters
record
interfaces
alarm limits
curves
basic configurations
IPPV default settings
PCV default settings
fresh gas delivery
85