Technical Data
Technical Data

Ambient conditions

In operation:
Temperature
Air pressure
Rel. humidity
CO
concentration of ambient air
2
In storage:
Temperature
Air pressure
Rel. humidity

Fresh gas

(electronic mixer)
Settings:
O
concentration
2
Fresh gas flow
O
flush
2

Ventilator

(electronically controlled, pneumatically
driven bellows in bottle ventilator)
Settings:
Pressure limit Pmax
for IPPV
for PCV
Tidal volume
Ventilation frequency
T
insp
T
exp
Insp/Exp. time ratio T
: T
I
E
Plateau time ratio T
: T
IP
I
Inspiratory flow (PCV)
Inspiratory flow (IPPV)
(cannot be set directly; depends on
the setting for V
, Freq. T
:T
and T
:T
)
T
I
E
IP
I
PEEP
10 to 35 °C
700 to 1060 hPa
20 to 80 % (no condensation)
300 to 800 ppm
–20 to 60 °C
sensor max. 50 °C
O
2
500 to 1100 hPa
<98 % (no condensation)
25 to 100% by vol., at least 25% by vol., 250 mL/min.
± 5% of set value
0 and 0.5 to 12 L/min
± 10% of set value or max. ±100 mL
>35 L/min
10 to 70 mbar
±10 % of setting or at least ±2 mbar
(PEEP+1) to 70 mbar
±10 % of setting or at least ±2 mbar
50 to 1400 mL
± 15% of the setting for inspiration flows ≥ 10 L/min.
6 to 60 per minute
0.2 to 6.7 seconds
0.33 to 8 seconds
1 : 4 to 2 : 1 with automatic ventilation
0 to 50 %
5 to 50 L/min ±10 %
Can be set to up to 75 L/min by DrägerService
3 to 75 L/min
0 to 20 mbar
±10 % of setting or at least ±2 mbar
Page 132