Setting the fresh gas flow
1 Press the softkey »L/min«.
The key field is highlighted against a light background.
Set and confirm the fresh gas flow by means of the
rotary knob.
Julian is fitted with an electronic O
system to avoid hypoxic gas mixtures. For fresh gas flow
settings below 1 L/min, the O
automatically increased to a value corresponding to an
O
flow of 250 mL/min. If this control system is
2
activated, the O
value is highlighted in addition to the
2
active softkey »L/min«.
Version with fresh gas failure detection (optional):
During operation, Julian checks that the bellows are
sufficiently full.
If the message „Fresh gas ? !!" appears:
Increase the fresh gas flow
The default settings valid whenever Julian is switched on
can also be modified; see "Setting default values" on
page 82.
Selecting the ventilation mode
MAN/SPONT ventilation mode
Choose between manual ventilation MAN and sponta-
neous breathing SPONT on the APL pressure-limiting
valve.
Manual ventilation
2 Set the lever of the APL pressure-limiting valve to
MAN.
3 Set the pressure limit = rotate lever.
Spontaneous breathing
4 Set the lever of the APL pressure-limiting valve to
SPONT.
The valve is open for a free spontaneous breathing
stroke regardless of the set pressure limit.
minimum dosing
2
concentration is
2
Setting the fresh gas flow
Selecting the ventilation mode
Julian
1
2
3
4
Operation
MAN/SPONT