Manual calibration

Julian calibrates its sensors automatically during
operation. The message CAL appears instead of the
current measured value during a calibration process.
Calibration of the O
sensor and zero calibration of the
2
other gas sensors can be repeated manually, for instance
if automatic calibration was unsuccessful.
Turn the rotary knob to select the »calibration«
column and press to confirm.
Select and confirm the appropriate line with the rotary
knob, for example:
»O
-Sensor
2
21 vol.%«
The calibration process is started when the selection is
confirmed. The clock symbol
relevant line, i.e. calibration is in progress.
When calibration is complete, a tick ( ) appears instead
of the clock symbol. A question mark (?) indicates that
calibration must be repeated because the previous
calibration was defective.
Other settings can already be made during the calibration
process.
U U U U
appears after the
IPPV
anesth. agent
alarms
PAW
selection
default
40
auto
man
CO
on
off
2
SpO
on
off
2
Isoflurane
20
HLM on
off
The anesthetic
agent is automati-
cally selected by
gas identification
0
Freshgas O
+ N
O
2
2
P
max
O
%
L/min
mbar
2
50
1.00
25
600 10 1:2 10
IPPV
PAW
O
40
21 Vol.%
gas sensor
20
0
Freshgas O
+ N
O
2
2
P
max
O
%
L/min
mbar
2
50
1.00
25
600 10 1:2 10
Monitoring
Configuring in operation
Manual calibration
calibration
settings
O
-Sensor
pulse tone
0 1 2 3 4
5
2
21 Vol.%
alarm sound
1 2 3
4
5 6 7 8 9
scale
adult
child
gas sensor
parameters
record
activate monitor defaults
V
Freq.
T
:T
T
:T
T
I
E
IP
I
mL
1/min
%
calibration
-sensor
2
V
Freq.
T
:T
T
:T
T
I
E
IP
I
mL
1/min
%
6 7 8 9
PEEP
mbar
0
PEEP
mbar
0
75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents