Before Using for the First Time
Charging the battery for emergency operation
When Julian is not in use
Before Using for the First Time
Fit the enclosed O
sensor,
2
page 117 "Replacing O
sensor"
2
Fit the flow sensor, page 106
Charging the battery for emergency
operation
Julian has a built-in uninterruptible power supply UPS
which maintains the power supply for 30 minutes in the
event of a mains failure, provided that the battery is
charged.
The switch-over to the battery-powered UPS is automatic
and is indicated on the screen by the message:
POWER FAIL!
The battery recharges automatically when the workstation
is plugged into the mains.
The battery must be charged for 10 hours before using
the Workstation for the first time.
Plug the mains plug of the Julian workstation into the
mains socket. The mains voltage must correspond to
the voltage specified on the nameplate.
N N N N
1 The green LED
lights up.
Leave Julian connected to the mains for 10 hours.
The workstation does not have to be switched on.
The devices connected to auxiliary power sockets will
not be powered by the UPS in the event of a power
failure.
When Julian is not in use
Charge battery at least every 4 weeks.
Allowing it to run down can lead to damage.
If Julian is out of use for an extended period:
Leave the workstation connected to the mains at all
times.