Alarm Info; Co 2 Alarms On /Off - Dräger Julian Instructions For Use Manual

Anaesthetic workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Monitoring
CO
alarms on/off
2

Alarm info

Alarm info
This function is used to list all active alarms / warnings in
order of priority.
Press and hold the softkey »alarm info«.
Display (example):
Active alarms:
FIO2 < < < < !!!
Active warning:
ETCO2 < < < < !!
The display only remains visible as long as the softkey is
pressed.
CO
alarms on /off
2
The alarm limits for etCO
2
monitoring can be switched off with the softkey
»CO2-al. on/off« in MAN/SPONT mode.
In this case, the symbol CO
measured end-expiratory CO
The CO
alarm limits are reactivated by pressing the
2
softkey again. The deactivated CO
reactivated automatically when changing to a different
ventilation mode.
The CO
alarm limits can also be activated and
2
deactivated under "Alarm limits" in the configuration
screen in all ventilation modes, see page 72.
National and European standards require a minimum
monitoring with alarm function. These standards may not
be met if the alarm function of the etCO
parameter is deactivated.
64
, insp. CO
and CO
apnoea
2
2
g
appears alongside the
2
concentration.
2
alarm limits are
2
monitoring
2
IPPV
Pleth
CO
2
40
0
PAW
20
0
Freshgas O
+ N
O
2
2
P
V
max
O
%
L/min
mbar
mL
2
50
1.00
25
600
MAN/SPONT
Pleth
CO
2
40
0
V
T
0.35
0
0.5
1
Volumeter --s
--
0
5
10
Start volumeter: Confirm
!
Freshgas O
+ N
O
2
2
fresh gas
O
%
L/min
internal
2
30
4.00
external
FIO2
!!!
FIO2
!!!
alarm
ETCO2
!!
limits
auto set
limits
alarm
info
list
curves
config.
Freq.
T
:T
T
:T
PEEP
T
I
E
IP
I
1/min
%
mbar
10
1:2
10
98
67
SpO
2
CO
38
2
etCO
freq
_
2
5.2
8
Ve
MV
Fi
Fet
40
36
O
2
0.8 0.6
Hal.
58
56
N
O
2
0
alarm
limits
CO
-al.
2
on /off
alarm
Info
list
curves
config.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents