Alarm mode for the heart-lung machine HLM

HLM alarm mode can be used independently of the
active ventilation mode for patient monitoring when using
the heart-lung machine.
In HLM alarm mode,
– all apnoea alarms are deactivated.
– the gas concentrations are measured continuously
regardless of the respiratory phase.
– the SpO
alarms are deactivated.
2
They are reactivated automatically when pulsations
are detected by the apparatus again after switching
off the HLM alarm mode.
Selecting HLM alarm mode
Press the softkey »config«.
Turn the rotary knob to select the column »alarms«
and press to confirm.
Turn the rotary knob to select the line »HLM on« and
press to confirm.
Display (example):
The lung is kept open with constant airway pressure.
HLM alarm mode can be switched off at any time:
Press the softkey »HLM off«.
In standby mode, the HLM alarm mode is automatically
deactivated.
MAN/SPONT HLM
CO