Configuring in Standby Mode

Manual calibration

Manual calibration
Julian calibrates its sensors automatically during opera-
tion. The message CAL appears instead of the current
measured value during a calibration process.
Calibration of the O
sensor and zero calibration of the
2
other gas sensors can be repeated manually, for instance
if automatic calibration was unsuccessful.
In standby / configuration mode:
Turn the rotary knob to select the »calibration«
column and press to confirm.
Select and confirm the appropriate line with the rotary
knob, for example:
»O
-sensor
2
21 vol.%«
The calibration process is started when the selection is
confirmed. The clock symbol Uappears after the relevant
line, i.e. calibration is in progress.
When calibration is complete, a tick ( ) appears instead
of the clock symbol. A question mark (?) indicates that
calibration must be repeated because the previous
calibration was defective.
To improve the measuring accuracy of high O
trations, an additional 100 vol% O
linearity test of the O
sensor can be performed under
2
the menu option »more«.
Other settings can already be made during the calibration
process.
100 Vol.% O
calibration
2
Use a separate O
source, e.g. O
2
O
from the Julian breathing system is not suitable, as
2
it still contains traces of anaesthetic gas.
Unscrew the sample line from the Y-piece and place it
in the continuous flow of the O
Turn the rotary knob to select the line
»O
-sensor
2
100 vol.%«
and press to confirm.
Allow the O
flow to continue flowing into the sample
2
line.
A tick ( ) appears after the field when calibration is
complete.
Screw the sample line back on to the Y-piece.
92
concen-
2
calibration and a
2
flowmeter.
2
source.
2
Standby / Configuration
anesth. agent
calibration
selection
O
-sensor
2
auto
man
21 vol.%
gas sensor
Isoflurane
more
The anesthetic
agent is automati-
cally selected by
gas identification
Standby / Configuration
calibration
O
-sensor
2
21 Vol.%
gas sensor
more
O
-sensor
2
100 Vol.%
linearity
test O
2
default settings
pulse tone
0 1 2 3 4
5
6 7 8 9
alarm sound
1 2 3
4
5 6 7 8 9
scale
adult
child
parameters
record
interfaces
alarm limits
curves
basic configurations
IPPV default settings
PCV default settings
fresh gas delivery