Monitoring
Configuring in operation
Setting the monitoring functions

Setting the measuring parameters

Select the line »parameters« with the rotary knob and
press to confirm.
The »parameters« menu appears in the left-hand field.
Display (example):
Switch SpO
measurement (optional) on/off:
2
Select the line »SpO
measurement« with the rotary
2
knob and press to confirm.
Select »on« or »off« with the rotary knob and press to
confirm.
Select the sample rate for gas measurement:
Select the line »sample rate« with the rotary knob and
press to confirm.
Select »60« or »200 mL/min« with the rotary knob
and press to confirm.
For paediatric use:
Select a sample rate of 200 mL/min in order to obtain
more accurate values on account of the high frequen-
cies and time constant of the measuring system.
Recirculate the measured gas, see page 109.
If the measured gas is not recirculated:
Increase the minute volume accordingly in order to
maintain the volume balance.
For adult use:
Select a sample rate of 60 or 200 mL/min.
Recommendation: select 200 mL/min and recirculate
the measured gas, see page 109.
Set CO
unit:
2
Select the line »CO
unit« with the rotary knob and
2
press to confirm. The last value selected is highlighted
by a grey background.
Select the required unit of measure by turning the
rotary knob and press to confirm.
70
IPPV
anesth. agent
alarms
calibration
PAW
40
SpO
-measure.
on
off
2
sample rate
60
200
mL/min
20
CO
unit
mmHg
kPa vol%
2
0
Menu for configuration of measurement parameters.
Freshgas O
+ N
O
2
2
P
V
max
O
%
L/min
mbar
mL
2
50
1.00
25
600 10 1:2 10
settings
puls tone
0 1 2 3 4
5
6 7 8 9
alarm sound
1 2 3
4
5 6 7 8 9
Scale
adult
child
parameter
record
activate monitor defaults
Freq.
T
:T
T
:T
PEEP
T
I
E
IP
I
1/min
%
mbar
0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents