Changing patients

To switch Julian to standby:
Press the Standby key
rotary knob.
The functions of the workstation are switched off.
The set alarm limits are cancelled and the default alarm
settings are valid again.
The default settings are loaded for dosing anaesthetic
gas and for the ventilation parameters.
If the ventilation hoses have been changed:
Perform a leakage test, page 48.

Changing soda lime

– If 2/3 of the absorber charge has turned violet.
– If the inspiratory CO
2
5 mmHg or more.
Press the Standby key
the rotary knob.
1 Raise the writing top.
2 Pull the catch and draw out the breathing system at
the same time.
3 Turn the absorber counterclockwise and pull it out
downwards.
Drain the used soda lime and dispose of with
domestic waste.
Fill the absorber with fresh soda lime up to the MAX
mark.
Fit the absorber to the breathing system from below
and turn it clockwise as far as it will go.
Push the breathing system inwards until it clicks into
place.
Fold down the writing top.
E
and confirm with the
concentration FiCO
equals
2
E
and confirm by pressing
Standby
Julian
SW-version 2.0
01. Nov. 98
MAN
Operation with
SPONT
Freshgas O
+N
O
2
2
O
L/min
%
2
30
4.00
MAX
3
Operation
Changing patients
Changing soda lime
leakage
test
delete
trend
config.
,
IPPV
or
PCV
1
2
47